5.CALIBRATION
45
5-7-13. Min.scale div.(FLOW)
This function specifies the minimum measurement division (scale interval) for belt
speed, feed rate, and feed ratio.
Operation
5-7-14. Z_error upper limit
This is the set value with which to monitor load ratio while performing automatic zero-
point adjustment using
key or control signal input.
Measurements cannot start while load ratio exceeds the set value. Also, measurements
will be aborted when the measured value exceed it during operation, generating a
ZERO ERR. MEAS. NG control message.
Operation
5-7-15. Zero point auto adjustment
Zero point auto-adjustment are triggered either by
key or control signal
(rising edge on the 28 pin). The zero point adjustment is performed in test mode, and
ZERO ERROR SET value is automatically updated.
Note that the measurement cannot start while load ratio exceeds Z_ERR upper limit.
Also, measurements will be aborted when the measured value exceed it during
operation. Exits from measurement caused by pressing
key or the arrival of
falling edge on 28-pin are also considered as abnormal exits, generating the ZERO
ERR. MEAS. NG control message. Abnormal ends do not update the ZERO ERROR
SET value: the value before automatic zero adjustment is retained.
Abnormal ends generate a ZERO ERR. MEAS. NG control message.
Range: 1 – 100 (3 digits, decimal point moveable)
MODE
FLOW CALIBRATION
→
→
MIN.SCALE DIV.(FLOW)
PAGE
→
Z_ERR
Range: 0 – 999.99%
(
5 digits, decimal place fixed
)
MODE
FLOW CALIBRATION
→
→
Z_ERR UPPER LIMIT
PAGE
→
Z_ERR
STOP