11.TROUBLESHOOTING
167
11-10. ANA Error (Analog IN/OUT interface required)
ANA ERR 1
Input signal was below the minimum value when AI Zero Set was defined. Make sure
the input signal falls in the allowable range and try to set zero again.
ANA ERR 2
Input signal was above the maximum value when AI Zero Set was defined. Make sure
the input signal falls in the allowable range and try to set zero again.
ANA ERR 3
Input signal was below the minimum value when AI Full Scale Set was defined. Make
sure the input signal falls in the allowable range and try to set full scale again.
ANA ERR 4
Input signal was above the maximum value when AI Full Scale Set was defined. Make
sure the input signal falls in the allowable range and try to set full scale again.
ANA ERR 6
Input signal was too small to secure 1/1000 resolution when AI Full Scale Set was
defined (stable scale graduation is impossible). Make sure that the magnitude of input
signal is correct, and try to set full scale again.
ANA ERR 7
When AI Full Scale was defined, the input signal had shifted to negative direction than
that used for AI Zero Set. Make sure that the magnitude of input signal is correct, and
try to set full scale again.
AI Zero Set input range:
Current: +3.2 mA - +4.8mA
Voltage: -0.5V - +0.5V
AI Full Scale Set input range:
Current: +19.2mA - +20.8mA
Voltage: +9.5V - +10.5V