CWD2005 DPC
Operation combustion calorimeter
chapter — page:
7—45
7.6.1.1.1.4 Main menu - Options - I/O - Digital inputs
Signal and zero position are list fields. These inputs can start a calibration or switch
a holding amplifier on or off. The inputs are queried once per second; i. e. a change
must be pending for at least one second in order to be detected by the system.
Signals
Zero position
Contact
CWD2005
Start calibration
high
open
Calibration starts
closed
No calibration
low
open
No calibration
closed
Calibration starts
Start measurement
high
open
Measurement starts
closed
Measurement stops
low
open
Measurement stops
closed
Measurement starts
Signals hold
high
open
mA hold
closed
mA online
low
open
mA online
closed
mA hold
Calibration cancel
high
open
Calibration cancellation
closed
No calibration cancellation
low
open
No calibration cancellation
closed
Calibration cancellation