CWD2005 DPC
chapter — page:
5—12
5.4.5 Process gas
DANGER
Risk of explosion!
Danger due to invalid line glands!
If improper line glands are used, the explosion protection cannot be ensured any-
more!
•
Only use line glands approved for the required ignition protection type!
•
Observe the technical data when selecting / using cable and line glands!
Danger due to open drilled holes or unused line glands!
If drilled holes are open or cable glands unused, the explosion protection cannot
be ensured anymore!
•
Always close open drilled holes and unused line glands with approved
sealing plugs!
•
Observe the technical data when selecting/replacing suitable sealing
plugs!
NOTE
•
The connecting parts need to be clean and free of residue. Impurities can en-
ter the calorimeter and cause incorrect measurements and/or damage.
•
The inlet pressure for the gas connections must not exceed the values speci-
fied in the technical data of the combustion calorimeter.
•
Each connection needs to be carefully checked for leaks. If there are any
leaks, the system will draw air, and the measurements will be incorrect.
•
Do not use sealing agents to seal the gas connections. Sealing agents can
distort measurements.
•
Only use suitable pipes.
ATTENTION
The process gas must be free of condensate and dust if the system has no gas
preparation system!
Summary of Contents for CWD2005 DPC
Page 1: ...Operating instructions Combustion calorimeter CWD2005 DPC Mai 2017 V2 01 ...
Page 12: ...CWD2005 DPC chapter page 2 2 ...
Page 28: ...CWD2005 DPC chapter page 4 6 ...
Page 114: ...CWD2005 DPC chapter page 9 2 ...
Page 116: ...CWD2005 DPC chapter page 10 2 ...
Page 119: ...CWD2005 DPC Annex chapter page 11 5 ...
Page 120: ...CWD2005 DPC Annex February 2015 V2 00 chapter page 11 6 ...