CWD2005 DPC
Maintenance and service
chapter — page:
8—3
DANGER
Risk of explosion!
If the pressure monitoring system is bridged for maintenance / service work, the
device can be operated permanently without compressed air purging and over-
pressure.
There is no protection against explosions!
•
Bridging only for service and maintenance work!
•
Bridging only by authorised trained staff!
•
Use suitable measures / warnings to prevent permanent operation!
Fig. 8.1:
Top hat rail X9, Service bridge
Item
No.
Designation
Function / activity
1
Service bridge
bridge pressure monitoring system, combustion calorimeter
can be used without protection system, Terminal 3L and 5L
1