CWD2005 DPC
chapter — page:
7—58
7.6.1.1.3.10
Main menu - Options - System - Settings - Hardware 1
This menu should only be modified after consulting the manufacturer.
It controls the limits of measuring devices over two or three ranges. The overlap is
optimised. For this purpose you need to specify functional parameters. The time
constants of the burners are specified.
The calibration of the values of the carrier gas is entered and the time of connection
is determined. In case of changing the carrier gas, the manufacturer can provide
the new data. They are entered in the screen mask shown below.
All values are tested according to customer's specifications before delivery.
The screen shows all possible configurations.
Range No.
indicates the number of installed measuring ranges. A maximum of 3
measuring ranges are possible. In this example, 3 measuring ranges are config-
ured.
Dia. WJet.
indicates the nozzle diameter in mm of the measuring range nozzle.
This specification is information only and is meant to help later in calculating the
gas volumes.