5.4.1 Setup site
The installation location of the system must meet the following requirements:
•
Observe the requirements acc. to the technical data, especially the ambient
conditions and the IP protection type.
•
Clean, closed room
•
Protected from direct sunlight
•
Insure a clean, sufficient amount of compressed / ambient air for undistorted
measurements
•
Ensure that the load-bearing capacity of the environment is sufficient
•
Design sufficient space on the sides for mounting and connecting lines
•
Install according to the requirements of explosion protection marking
WARNING
Leaking process gas needs to be discharged by the operator into a safe
environment!
NOTE
The compressed air must be dry and free from dust and oil!
Humid, dirty air will distort the measurement results of the analyser.