CWD2005 DPC
chapter — page:
5—20
1.
Terminal box for power supply
Fig. 5.7:
Electric connection system – power supply X16
Assignment see provided wiring diagram! Connect the system via terminals L1, N,
PE to the voltage supply according to national/internal regulations.
X16.16.
Connection L1
X16.18.
Connection N1
X16.20.
Connection PE
The power supply must be designed according to the power consumption, refer to
type plate. The circuit must be protected by an external circuit breaker according
to national requirements, ratings according to type plate und technical data.
Connection Pressure Switch via Relay X8.K2, connection to terminal X17 installed,
refer to Fig. 5.9
X16.2.
COM
X16.4.
NO
X16.6.
NC
Terminal block X16
Terminal block X7
in terminal box (Ex e) for power supply
in purge enclosure
Customer
Union Instr.
Cable
Union Instr.
Relay
X16.2
Relay X8.K2
COM
X16.1
7G1,5.3
X7.K1.11
(signal
X16.4
NO
X16.3
7G1,5.4
X7.K1.14
overpressure)
X16.6
NC
X16.5
7G1,5.5
X7.K1.12
[…]
[…]
[…]
[…]
S1.13
X16.10
switched L1
L1
X16.9
7G1,5.1
X7.2L
S1.23
X16.12
switched N
N
X16.11
7G1,5.2
X7.2N
from PE
X16.14
PE
PE
X16.13
7G1,5.PE
X7.2PE
Power
X16.16
Power supply
L1
X16.15
MS 1,5 B
S1.14
switched L1
supply
X16.18
N
X16.17
MS 1,5 BU
S1.24
switched N
X16.20
PE
X16.19
MS 1,5 PE
PE
to PE bar
7G1,5.6
not connected
1
Summary of Contents for CWD2005 DPC
Page 1: ...Operating instructions Combustion calorimeter CWD2005 DPC Mai 2017 V2 01 ...
Page 12: ...CWD2005 DPC chapter page 2 2 ...
Page 28: ...CWD2005 DPC chapter page 4 6 ...
Page 114: ...CWD2005 DPC chapter page 9 2 ...
Page 116: ...CWD2005 DPC chapter page 10 2 ...
Page 119: ...CWD2005 DPC Annex chapter page 11 5 ...
Page 120: ...CWD2005 DPC Annex February 2015 V2 00 chapter page 11 6 ...