CWD2005 DPC
chapter — page:
5—18
5.4.11 Electrical connection
DANGER
Danger from electrical shock!
Only a trained electrician shall connect the power supply and electrical signal
wires!
Only a trained electrician may modify the electrical equipment of the system in
accordance with the relevant guidelines!
When the system has been opened, the parts identified by the adjacent symbol
may still be live even when the master switch has been turned off. If necessary,
disconnect system from the voltage mains!
The electric line to the power supply must meet the requirements for operation in
potentially explosive atmospheres, e.g. IEC/EN 60079-14, as well as national reg-
ulations.
The cable gland must meet the requirements of EX zones and be matched to the
cable and the threaded hole in the housing. Observe the operating instructions of
the cable glands!
Integrate all blank, not energised metal parts independent of the operating voltage
into the protective ground system.
The outer protective ground connection at the system is designed for cable lugs.
Avoid loosening of the cable, lay the cable close to the housing.
5.4.12 Electrical interfaces
WARNING
Untrained personnel starting the system may endanger people and equipment.
Only trained service technicians may start up the analyser.
DANGER
Danger of explosion!
Not used holes for lines must be closed with stopper plugs. Tighten stopper plugs
with tools. Use stopper plugs supplied by the terminal box manufacturer!
Summary of Contents for CWD2005 DPC
Page 1: ...Operating instructions Combustion calorimeter CWD2005 DPC Mai 2017 V2 01 ...
Page 12: ...CWD2005 DPC chapter page 2 2 ...
Page 28: ...CWD2005 DPC chapter page 4 6 ...
Page 114: ...CWD2005 DPC chapter page 9 2 ...
Page 116: ...CWD2005 DPC chapter page 10 2 ...
Page 119: ...CWD2005 DPC Annex chapter page 11 5 ...
Page 120: ...CWD2005 DPC Annex February 2015 V2 00 chapter page 11 6 ...