CWD2005 DPC
chapter — page:
8—6
NOTE
The following overview only describes the checks / work in regard to the protec-
tive housing and the functions for explosion protection!
Inspection
Interval
(recommended)
Weekly inspection
Compressed air inlet free
weekly, or when
required
Pressure indicator at protective cabinet readable and plausible
weekly, or when
required
Flame arrester consistent and not soiled
weekly, or when
required
Exhaust system free
weekly, or when
required
weekly, or when
required
Quarterly inspection
Process gas analyser calibrated according to manufacturer's speci-
fications
every 3 months, lat-
est annually or when
required