51
NT UF 841 CP GB (d) / June 2015
7) Logger variables
For each of the variables of the logger, one of the following functions can be selected:
•
AVG: average value over the recording period,
•
MIN:
minimum value over the recording period,
•
MAX:
maximum value over the recording period.
8) Log on variation
The variation mode of the logger is used to reduce the recording period for the data in the logger.
To do this, define the maximum variation percentage with regard to the previous recording. If the variation
measured is greater than this maximum, recording is instantly triggered (no more than once per second).
Comment:
A percentage equal to 0% disables the associated value.
The following figure shows the reduction in the period of the logger when the flow experiences a variation:
The vertical lines (blue) show when logger recording is triggered. If the variation of the flow is greater than
the configured threshold, the logger forces a recording.
Time (s)
Flow
(m
3
/s)
Flow measurement
Logger
Summary of Contents for Uf 841
Page 4: ...4 NT UF 841 CP GB d June 2015 CHAPTER 1 OVERVIEW...
Page 12: ...12 NT UF 841 CP GB d June 2015 CHAPTER 2 SAFETY INSTRUCTIONS...
Page 17: ...17 NT UF 841 CP GB d June 2015 CHAPTER 3 INSTALLATION AND WIRING...
Page 25: ...25 NT UF 841 CP GB d June 2015 CHAPTER 4 INSTALLING A MEASURING POINT...
Page 27: ...27 NT UF 841 CP GB d June 2015...
Page 35: ...35 NT UF 841 CP GB d June 2015 CHAPTER 5 USING AND CONFIGURING THE Uf 841...
Page 64: ...64 NT UF 841 CP GB d June 2015 CHAPTER 6 FUNCTION ENGINE...
Page 68: ...68 NT UF 841 CP GB d June 2015 CHAPTER 7 PC SOFTWARE...
Page 82: ...82 NT UF 841 CP GB d June 2015 Appendix I Characteristic echo signals...
Page 85: ...85 NT UF 841 CP GB d June 2015 Appendix II Speed of sound in water...