Page 92
Message /
Symbol /
Problem
Failure Description
Corrective Action
stage.
High
Pressure
This message is displayed if
Chamber Pressure raises 4.2
psi-29 kPa above sterilization
pressure (134°C = 304 kPa,
121°C = 205 kPa) for 2
seconds during the
sterilization stage.
Perform a new cycle.
High
Pressure
(Ending)
This message is displayed if
the system cannot reach
atmospheric pressure ±
10kPa during the ending
stage.
Perform a new cycle.
High
Pressure
(Exhaust)
This message is displayed if
the system cannot reach
preset pressure within 10
minutes from the beginning of
the exhaust stage.
Perform a new cycle.
Pressure
Time Error
This message is displayed if
the
system cannot reach the
required pressure conditions
in the chamber, after preset
time, during the air removal
stage.
Verify that the autoclave
is not overloaded.
RTC Error -
Please Set
Current Date
and Time
This message is displayed in
order to set the date and the
time.
Set Current Date and
Time.
If the problem persists,
call the technician.
Time Error
This message is displayed if
the real time clock is faulty.
Call the technician.
Door is open
(During the
cycle)
This message is displayed
when the door is open:
During the cycle.
Close the door to
perform a new cycle.
Canceled By
User
This message is displayed
after the START/STOP key is
pressed and cycle aborted.
Wait until "cycle failed
–
canceled by user" or
"cycle end
– canceled by
user" is displayed.
Perform a new cycle.
Cycle
Failed
This message and symbol
are displayed if an error
occurs before sterilization
cycle is completed.
Perform a new cycle.
Summary of Contents for LabSci 11-V
Page 19: ...Page 18 8 Right service door lock...
Page 22: ...Page 21 10 Drain outlet...
Page 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Page 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Page 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Page 79: ...Page 78...