Page 41
•
Heating:
The chamber is heated by actuation of electrical heaters
until the sterilization temperature is reached.
•
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
•
Cooling:
Forced cooling to the required end temperature, chamber
pressure remains at approx. 300 kPa and the cooling valve is
opened.
•
Fast Exhaust:
The steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
Drying:
N/A
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.7
Program 7: Bio Hazard 1 (BH, bio-hazard models
only)
All exhaust from the chamber before completion of the
Sterilization stage is performed through the bio-hazard filter.
For
instruments when the manufacturer recommends autoclaving at
temperatures of 134
C (273
F) for 30 minutes. Drying stage is
available for PV (pre-vacuum) models only.
Nominal parameters default settings
•
Sterilization temperature: 134
C (273
F)
•
Sterilization time: 30 minutes
•
Drying time: 1 minute. (PV
– pre-vacuum models only).
Operations sequence: