Page 85
For detailed loading instructions, see sec. 9 (Preparation before
sterilization)
10.11.3
Unloading
On completion of the cycle, take out the load immediately from the
sterilizer. Do not remove the load from the basket until its temperature
reduces to the room temperature. Let the load cool down in an area
without air movement (air conditioning, etc.) and with minimum people
passing by to avoid possibility of touching the hot load. Do not touch
the hot load since hot load absorbs moister and, therefore, may absorb
bacteria from your hand. Do not transfer hot load to metal shelves for
cooling. Perform a visual inspection to ascertain that sterilizing
indicators have made the required color change, and that the load is
dry.
The load shall be rejected if:
a.
The package has been compressed.
b.
The package is torn.
c.
The load is suspected to be wet.
d.
The load fell on the floor.
e.
Condensed drops can be detected on the load.
To avoid injuries, use heat resistant gloves while unloading
the autoclave.