Page 25
flow is stopped automatically, process is completed, and it is possible
to open the door and take out the sterilized goods from the chamber.
3.17.4
Exhaust
When the timed sterilization cycle is complete, the unit enters the
exhaust stage, provided that a program other than the liquid program
was selected. The steam is exhausted from the chamber, bringing the
internal pressure down to atmospheric pressure.