Page 13
Rest of heavy metals
except iron,
cadmium, lead
≤ 0.1 mg/l
≤ 0.1 mg/l
Chloride (Cl)
≤ 2 mg/l
≤ 0.1 mg/l
Phosphate
≤ 0.5 mg/l
≤ 0.1 mg/l
Conductivity (at
20°C)
15 µs/cm
≤ 3 µs/cm
pH value
5 to 7.5
5 to 7
Hardness
≤ 0.02 mmol/l
≤ 0.02 mmol/l
Appearance
Colorless, clean, without sediments
Note:
The condensate is produced from steam taken from the empty
sterilizer chamber.
Compliance with the above data should be tested in accordance with
acknowledged analytical methods, by an authorized laboratory.
Attention:
The use of water for autoclaves that do not comply with the table
above may have severe impact on the working life of the sterilizer
and can inval
idate the manufacturer’s guarantee.
Use only deionized water, having a maximum conductivity of 15
µs/cm. Conductivity greater than 15 µs/cm may cause failures.
Tap water supply
The range of hardness value 0.7-2.0 mmol/l (70- 200 mg/l CaCO
3
)
The use of soft water is strictly forbidden!
Please consult a water specialist!
3.7.1
Drain Cooling
The feed water supplied to the drain cooling must meet the following
requirements:
•
Hardness: 0.7 - 0.2 mmol/l.
•
Water temperature shall not exceed 15°C (59°F).
Summary of Contents for LabSci 11-V
Page 19: ...Page 18 8 Right service door lock...
Page 22: ...Page 21 10 Drain outlet...
Page 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Page 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Page 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Page 79: ...Page 78...