Page 79
10.
Operating Instructions
10.1
Draining and filling the water reservoir (WR only).
Draining
CAUTION!
Before starting, make sure that the electric cord is disconnected
and there is no pressure in the autoclave.
The drain valve is located on the bottom of the water reservoir (See
3.14 .
To drain the reservoir into a bucket:
1.
Open the drain valve.
2.
Fully drain the reservoir.
3.
Close the drain valve.
Note:
Drain the water from the mineral free water reservoir once a
week, and refill it with fresh mineral-free water or distilled water.
Filling
1.
Connect the electric cord to power source.
2.
Pour distilled water through a water inlet near the chamber into the
reservoir using a carafe so that the glass tube fills 4-5 cm below its
upper end as shown by the yellow float (see the figures below). The
water quantity is approx. 11.7 liters.