Page 40
•
Cooling:
Forced cooling to the required end temperature, chamber
pressure remains at approx. 300 kPa and the cooling valve is
opened.
•
Fast Exhaust:
The steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
Drying:
N/A
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.6
Program 6: Liquid B
–
Waste Cooling (C
–
cooling
models only)
For liquids when the manufacturer recommends autoclaving at
temperatures of 121ºC (250
F) for 30 minutes, such as liquid waste.
Cautions!
Both PT100 temperature sensors must be inside the bottles.
For proper sterilization, fill the bottles with approximately the
same amount of liquid.
Nominal parameters default settings
•
Sterilization temperature: 121
C (150
F).
•
Sterilization time: 30 minutes.
Operations sequence:
•
(PV, pre-vacuum models only)
Pulse low/Pulse high:
At one
pulse it will build vacuum down to 25 kPa.
•
Water Inlet:
Water enters the chamber