Page 15
3.9
Construction
Chamber and door material
Stainless steel
Outer Cabinet
Stainless Steel
Chamber insulation
Fiberglass with reinforced material
3.10
Utilities
Power Supply
Power
1 Ph, 230V/50/60Hz
Recommended
Circuit Breaker
16A
Compressed Air (LVC only)
1/2" 3 Bar (44-58 psi)
Tap water
1/2", 2-6 Bar (29-44 psi)
Mineral free water
1/2", 2-3 Bar (29-44 psi)
Drain
2" Minimum Withstanding temp. of 80 M
(17676)
Attention!
•
A switch or circuit-breaker must be included in the building
installation. This switch or circuit-breaker shall be in close
proximity to the equipment, within easy reach of the operator; and
marked as the disconnecting device for the equipment.
•
The electrical net must be protected with a current leakage safety
relay.
•
The electrical network must comply with local rules or
regulations.
•
Verify that there is an easy access to the main power switch, to
the water cut-off valve and to the current leakage safety relay.
•
Make sure while placing the autoclave, to leave space around the
machine, to give the technician access to service the machine.
•
All water connections to autoclave must be performed through
"BACK FLOW PREVENTION SYSTEM" only, as per IEC 61770.
Summary of Contents for LabSci 11-V
Page 19: ...Page 18 8 Right service door lock...
Page 22: ...Page 21 10 Drain outlet...
Page 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Page 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Page 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Page 79: ...Page 78...