Page 42
•
(PV, pre-vacuum models only)
Pulse low/Pulse high:
At one
pulse it will build vacuum down to 25 kPa.
All exhaust from the
chamber is performed through the bio-hazard filter.
•
Water Inlet:
Water enters the chamber
•
Heating:
The chamber is heated by actuation of electrical heaters
until the sterilization temperature is reached.
•
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
•
Cooling:
N/A
•
Fast Exhaust:
The steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
Note:
If the cycle fails, fast exhaust is performed through the bio-hazard
filter. Some water may remain in the chamber after a failed cycle. To
remove such water, press Start and repeat the cycle. If the problem
persists, call for service.
•
(PV, pre-vacuum models only)
Drying:
Vacuum is built up and the
air pump is working.
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.8
Program 8: Bio Hazard 2 (BH, bio-hazard models
only)
All exhaust from the chamber before completion of the
Sterilization stage is performed through the bio-hazard filter
. For
instruments when the manufacturer recommends autoclaving at
temperatures of 121
C (250
F) for 45 minutes. Drying stage is
available for PV (pre-vacuum) models only.
Nominal parameters default settings
•
Sterilization temperature: 121
C (250
F)
Summary of Contents for LabSci 11-V
Page 19: ...Page 18 8 Right service door lock...
Page 22: ...Page 21 10 Drain outlet...
Page 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Page 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Page 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Page 79: ...Page 78...