Program 11: Warm-Up
Pre-warming the chamber at 80
C (176
F) without drying.
Caution!
This is not a sterilization program!
Nominal parameters default settings
•
Warm-up temperature: 80
C (176
F).
•
Warm-up time: 20 minutes.
Operations sequence:
•
Pulse low/Pulse high:
N/A
•
Heating:
Water enters the chamber and is heated by actuation of
electrical heaters until the warm-up temperature) is reached.
•
Warm up:
Warm up temperature is maintained constant for the
warm-up time.
•
Sterilization:
N/A.
•
Cooling:
N/A.
•
Fast Exhaust:
The steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
Drying:
N/A.
Note:
Some water may remain in the chamber.