Page 12
The baskets are made of stainless-steel wire and have a handle. The
basket allows the operator to load a large quantity of materials into the
chamber.
•
Stainless steel containers - Different size containers are available for
these units.
The containers are designed for sterilizing waste material. The
containers have vent holes along the upper rim.
3.5
Environment Emission Information
A.
Peak sound level generated by the sterilizer is « 70 / dBA with a back-
sound level of 60 dBA.
B.
Total heat transmitted by the sterilizer is < 240 W/h
3.6
Electrical Data
3.7
Water quality
The distilled or mineral-free water supply shall be according to the table
below:
Suggested maximum limits of contaminants in water for steam
sterilization per EN13060
Substance
Feed Water
Condensate
Evaporate residue
≤ 10 mg/l
≤ 1.0 mg/l
Silicate (SiO
2
)
≤ 1 mg/l
≤ 0.1 mg/l
Iron
≤ 0.2mg/l
≤ 0.1mg/l
Cadmium
≤ 0.005 mg/l
≤ 0.005 mg/l
Lead
≤ 0.05 mg/l
≤ 0.05 mg/l
Property
Value
Total Power
2400W (2*1200W)
Voltage
230VAC/1Ph
Amperage
10A
Protection against electrical shock
Class I (IEC 61010-1)
Mains supply fluctuation
+/- 10%
Degree of protection by enclosure
IP31
Summary of Contents for LabSci 11-V
Page 19: ...Page 18 8 Right service door lock...
Page 22: ...Page 21 10 Drain outlet...
Page 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Page 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Page 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Page 79: ...Page 78...