Software Features
SeaKing Profiling Sonar
0374-SOM-00002, Issue: 04
44
© Tritech International Ltd.
This will launch the output configuration dialog named
Slot 2, Output 1
:
Within this dialog it will be necessary to select the
Channel
which was
previously configured and then click on
Gate Alarm On
to enable the alarm
to be transmitted over the COM port (when the alarm is activated in Seanet
Pro an ASCII string will be sent out).
Note
Do not set
Continuous Data Output
unless the full data string
from the Profiler is also to be sent out on the same COM port as
the alarm message.
5.2.4. Remote Alarm String Format
The transmitted Alarm message is a serial ASCII string with the following
format:
%M0026022530120707620000100008000012<CR><LF>
%M0026022530 Message Header {&M = Fixed Header String, 0026 =
Number of bytes in message (in hexadecimal), 02 =
RemV4 slot number, 25 = fixed device type of Profiler (in
hexadecimal), 3 = Ignore, 0 = Ignore}.
12070762
Time of Alarm (hhmmssdd = Hours,Min,Sec,DeciSec).
0
Alarm Mode selected (0 = Outside Gate, 1 = Inside Gate).
00010
Point Threshold (number of points used in alarm threshold,
as configured in Seanet Pro).
00080
Total Number of points in profile.
00012
Number of Points (from the total) that are outside the gating
zone.