Operation
SeaKing Profiling Sonar
0374-SOM-00002, Issue: 04
23
© Tritech International Ltd.
5. Wait until the download has completed and until the table has been rebuilt.
The node status should change from
Update
to
OK
to indicate that the
download has been completed.
To Change a Node Number
Note
The different types of device are programmed with different ranges
of node numbers and these devices will not operate if a node value
is selected which is out of this range. For a full list of ranges for
different
Tritech International Ltd
devices please refer to the Seanet
Pro manual (document reference: 0706-SOM-00001).
Node ranges:
20 - 29
SeaKing DF Profiler
252
SCUv5 or SeaHub
254
SKIM-100 Interface Box (obsolete)
255
ARCNET Interface Card (ISA or PCI AIFV4 card in a PC)
Changing node numbers is useful for running a Profiler application when 2
heads are connected that originally have the same node number (factory
supplied pairs should already be set up as master and slave). To reprogram a
node the Profiler heads to have the node number re-programmed should be
connected by itself and then follow the procedures outlined here to change
its node number.
!
Caution
Only re-program nodes over short lengths of test cable (1m lengths
are available from
Tritech International Ltd
if not already supplied
with the sonar head at time of purchase). Longer cables increase
the risk of the data becoming corrupted in transit and could result
in the sonar head becoming unusable after the update.
1. Open Seanet Setup and wait for a few seconds for the nodes to build.