Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
20 mA value
One of the following options is
selected in the Current span
parameter :
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
Enter 20 mA value.
Signed floating-point
number
Depends on country
and nominal
diameter
Fixed current
The Fixed current option is
selected in the Current span
parameter .
Defines the fixed output
current.
3.59 to 22.5 mA
4 mA
Failure mode
A process variable is selected
in the Assign current output
parameter and one of the
following options is
selected in the Current span
parameter :
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
Define output behavior in
alarm condition.
• Min.
• Max.
• Last valid value
• Actual value
• Defined value
Max.
Failure current
The Defined value option is
selected in the Failure mode
parameter.
Enter current output value in
alarm condition.
3.59 to 22.5 mA
22.5 mA
*
Visibility depends on order options or device settings
Damping output 1
‒
Set the reaction time of the
output signal of the current
output to fluctuations in the
measured value.
0 to 999.9 s
1 s