Flow conditioner
If the inlet runs cannot be observed, the use of a flow conditioner is recommended.
The flow conditioner is fitted between two pipe flanges and centered by the mounting
bolts. Generally this reduces the inlet run needed to 10 × DN or 13 × DN with full accuracy.
1
3
4
2
5
6
7
8
9
h
Minimum inlet and outlet runs with various flow obstructions
1
2
3
4
5
6
7
8
9
Concentric reducer
Eccentric reducer
Single elbow (90°elbow)
Double elbow (2 × 90°elbows, on one plane)
Double elbow 3D (2 × 90° elbows, not on one plane)
T-piece
Expansion
Control valve
Combination pipe (Double elbow 3D (2 × 90° elbows, opposite, not on one lane) + reducer, etc.)
5 × DN 8 × DN
5 × DN 8 × DN
2 × DN 8 × DN
5 × DN 8 × DN
5 × DN
5 × DN
5 × DN
5 × DN
2 × DN 8 × DN
5 × DN
2 × DN 8 × DN
5 × DN
5 × DN 8 × DN
5 × DN
5 × DN 8 × DN
5 × DN
5 × h 8 × DN
5 × DN
2 × DN
+
Flow Conditioner
Flow Conditioner
Flow Conditioner
Flow Conditioner
Flow Conditioner
Flow Conditioner
Flow Conditioner
Flow Conditioner
h: differece in
expansion
Flow Conditioner