EF200F-C
Installation
33
6.3
Post-installation check
Is the device undamaged (visual inspection)?
Does the measuring device conform to the measuring point specifications?
For example:
• Process temperature → See 16.9
• Process pressure (refer to the section on "Pressure-temperature ratings" in the "Technical
Information" document → See 16.15)
• Ambient temperature
• Measuring range → See 16.3
Has the correct orientation for the sensor been selected? → See 6.1.1
• According to sensor type
• According to medium temperature
• According to medium properties (outgassing, with entrained solids)
Does the arrow on the sensor nameplate match the direction of flow of the fluid through the
piping? → See 6.1.1
Are the measuring point identification and labeling correct (visual inspection)?
Is the device adequately protected against precipitation and direct sunlight?
Are the securing screw and securing clamp tightened securely?
Has the maximum permitted insulation height been observed?
• Has the pressure range been observed? → See 16.9
• Was the correct orientation selected? → See 6.1.1
• Is the pressure unit mounted correctly? → See 6.2.4
• Have the pressure gauge valve and the siphon with pressure sensor been mounted using
the prescribed seal and the specified torque → See 6.2.4
▢
▢
▢
▢
▢
▢
▢
▢
▢