5
•
Installation
Page 76
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
2.
If the column compartment is located to the right of the detector
Route the capillary through the bottom recess in the partition panel.
Make sure that you use the suitable partition panel plug for the
capillary.
TIP
For capillaries with small outer diameter, the plug with slit secures
the capillary properly. For capillaries that do not fit in the plug with slit,
such as an insulated inlet capillary, use the rotating plug.
3. Connect the inlet capillary to the flow cell inlet port (
IN
).
5.8.5.2
Connecting the Detector Waste Line
Preparations
1. See
Flow Connections to the Flow Cell (
.
2. Make sure you have the correct detector waste line for your flow
cell (see
).
Follow these steps
Figure 23: Connecting the detector waste line
1. Connect the waste line to the flow cell outlet (OUT) (1).
2. Route the waste line through the top recess (2) in the partition
panel.
3. Route the waste line through the tubing guides of the system
modules below the detector to the Vanquish system base (3).