Installation
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
Page 63
Follow these steps
1. Open the device doors.
2. Loosen the two screws of the flow cell cover, remove the cover and
keep the cover to close the flow cell opening after this procedure.
CAUTION—UV Radiation
The deuterium lamp emits UV radiation that is harmful to eyes and skin.
Do not look and do not reach into the flow cell opening while no flow
cell or diagnostic cell is installed and the UV lamp is turned on.
3. Hold the flow cell with the flow cell body to the left, firmly squeeze
the handle of the flow cell and insert it straight into the flow cell
opening. Release the handle. If the handle clicks, the cell is properly
mounted. If the flow cell does not click, move the flow cell slightly
just until the flow cell clicks into place.
After the flow cell has been installed, the firmware detects the flow
cell.
Figure 11: Inserting the flow cell (here: flow cell with connection unit)