7
•
Maintenance and Service
Page 154
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
7.14.2 Returning the Slide-In Module
Preparation
If not yet done:
Remove the slide-in module from the enclosure. See
Removing the Slide-In Module (
Follow these steps
1. Install the slide-in module to the enclosure of the replacement
device. Follow the steps for inserting the slide-in module in the
enclosure and tightening the captive screws in
.
2. Follow the instructions in
.
NOTICE
Shipping the slide-in module improperly leads to damage to the device.
Always ship the slide-in module as described in this operating manual.
CAUTION—Possible Contamination
Hazardous substances may have contaminated the device during
operation and may cause personal injury to service personnel.
•
Decontaminate all parts of the device that you want to return for
repair.
•
Fill in and sign the Health and Safety Form. Thermo Fisher Scientific
refuses to accept devices for repair if the Health and Safety Form is
missing, incompletely filled in, or unsigned.
7.14.3 Installing the Slide-In Module
CAUTION—Heavy Load, Bulky Device
The device is too heavy or bulky for one person alone to handle safely.
To avoid personal injury or damage to the device, observe the following
guidelines:
•
Physical handling of the device, including lifting or moving, requires
a team effort of two persons.
•
A team effort is in particular required when lifting the device into
the system stack or when removing it.
Tools required
Screwdriver, Torx T20