![ThermoFisher Scientific QuantStudio 12K Flex OpenArray AccuFill System User Manual Download Page 68](http://html1.mh-extra.com/html/thermofisher-scientific/quantstudio-12k-flex-openarray-accufill-system/quantstudio-12k-flex-openarray-accufill-system_user-manual_1105600068.webp)
Components (unpackaged)
Ensure that the installation site bench space can accommodate the dimensions and support the
weights of the components.
Component
Height
Length (depth)
Width
Weight
Instrument
50 cm (20 in.)
70 cm (27.5 in.)
80 cm (31.5 in.)
80 kg (176 lbs)
Computer
41.1 cm (16.2 in.)
43.8 cm (17.2 in.)
18.7 cm (7.4 in.)
14.3 kg (31.5 lbs)
Monitor (with stand)
38.1 cm (15.0 in.)
14.0 cm (5.5 in.)
37.5 cm (14.8 in.)
4.6 kg (10.2 lbs)
CAUTION!
PHYSICAL INJURY HAZARD.
Do not attempt to lift or move the instrument without
the assistance of others, the use of appropriate moving equipment, and proper lifting techniques.
Improper lifting can cause painful and permanent back injury. Depending on the weight, moving or
lifting an instrument may require two or more people.
Configured system dimensions and connections
A typical layout is shown in Figure 30
The components require a width of 1.8 m (6 ft).
Figure 30 Typical layout of the QuantStudio
™
12K Flex OpenArray
™
AccuFill
™
System
The dimensions of the whole system are shown in Figure 31
. For the dimensions of each
component in the system, see “Components (unpackaged)”
Appendix C
Site requirements
Configured system dimensions and connections