View the instrument log
Click
Service
4
Instrument Log
.
The instrument log is displayed.
Copy or export the instrument log
View the instrument log (see “View the instrument log”
1.
Click
Copy all
to copy the instrument log
The instrument log is copied to the computer clipboard and can be pasted into a text editor.
2.
Click
Export
, then navigate to a folder to export the instrument log.
The instrument log is saved as a CSV file.
Perform a system self-test (manual start)
Clear the instrument deck, empty the waste bin and place it on the instrument deck, then close the
instrument door.
The self-test checks the following items:
• Connection to the instrument
• Motion of the robotic tip handler on each of the three axes
• Tip ejection
1.
Click
Service
4
Self-Test
.
2.
In the
System self-test
dialog box, click
Start test
.
The system performs a self-test. The status of the self-test is displayed in the dialog box.
Appendix D
Manage the service settings
View the instrument log
D
84
QuantStudio
™
12K Flex OpenArray
™
AccuFill
™
System User Guide