• The system is within 2 m (6.6 ft) of a standard electrical outlet.
• There is sufficient space for the power cord to be easily unplugged or disconnected from the
system.
• The location is not near heaters, cooling ducts, or in direct sunlight. Fluctuations between day
temperatures and night temperatures can cause instrument instability.
• The location is not near a dust source, for example, nearby construction.
1
2
3
1
Fan
2
Power inlet receptacle
3
Computer cable receptacle
Environmental requirements
Condition
Acceptable range
Installation site
Indoor use only
Electromagnetic
interference
Do not use this device in close proximity to sources of strong electromagnetic radiation (for
example, unshielded intentional RF sources). Strong electromagnetic radiation may interfere
with the proper operation of the device.
This equipment has been designed and tested to CISPR 11 Class A. In a domestic
environment it may cause radio interference. You may need to take measures to mitigate
the interference.
Altitude
Located between sea level and 2,012 m (6,600 ft.) above sea level
Humidity (instrument
and computer)
Operation: 80% relative humidity for temperatures up to 31°C, decreasing linearly to 50%
relative humidity for 40°C
Temperature
(instrument and
computer)
15°C to 30°C (60°F to 85°F)
Note:
The room temperature must not fluctuate more than 2°C over a 2-hour period.