![ThermoFisher Scientific QuantStudio 12K Flex OpenArray AccuFill System User Manual Download Page 79](http://html1.mh-extra.com/html/thermofisher-scientific/quantstudio-12k-flex-openarray-accufill-system/quantstudio-12k-flex-openarray-accufill-system_user-manual_1105600079.webp)
3.
Select or deselect the
Interactive Test
checkbox.
If the checkbox is selected, the test includes a step to manually open and close the door.
If the checkbox is not selected, the test is completed without the door being manually opened and
closed.
4.
Enter a number in the
Run 1 of
field.
The number specified in this field is the total number of times that the instrument performs the
door lock test sequentially.
5.
Click
Start test
.
The status of each task in the door lock test is displayed.
If more than one run was specified, the number in the
Run 1 of
field is updated, for example,
Run
2 of
,
Run 3 of
.
6.
(For interactive tests only)
Follow the prompts in the
Door Lock Test
dialog box.
If the test fails, contact Service.
Perform a motion test
The motion test moves the robotic tip handler across the different axes.
• X-axis—Left and right
• Y-axis—Forward and back
• Z-axis—Up and down
1.
Click
Service
4
Diagnostics
.
2.
Select the
Motion Test
tab.
3.
Select the axes to test.
•
X Axis
•
Y Axis
•
Z Axis
4.
Enter a number in the
Run 1 of
field.
The number specified in this field is the total number of times that the instrument performs the
motion test sequentially.
5.
Click
Start test
.
The status of each task in the motion test is displayed.
If more than one run was specified, the number in the
Run 1 of
field is updated, for example,
Run
2 of
,
Run 3 of
.
If the test fails, contact Service.
Appendix D
Manage the service settings
Perform diagnostics tests
D
QuantStudio
™
12K Flex OpenArray
™
AccuFill
™
System User Guide
79