5.
Click
Done
.
The run does not proceed under any of the following conditions:
• The waste bin is not in position
• The sample plate is not in position
• The OpenArray
™
Plates are not in position
• There are more OpenArray
™
Plates on the instrument deck than are defined in the experiment
setup
• The instrument door is open
The instrument will proceed to load the next OpenArray
™
Plate.
6.
Repeat step 1 to step 5 for each OpenArray
™
Plate to be loaded.
After all of the OpenArray
™
Plates have been loaded, the
Deck
screen displays
Run completed
successfully. Empty the waste bin before performing another run.
If the run was started with a TPF or SPF file, a loaded TPF or SPF file is generated for each OpenArray
™
Plate. The loaded TPF or SPF file corresponds to the original TPF or SPF file that was imported for
the run. The files are exported to the folder that was designed in the
Preferences
).
If the run was started without a TPF or SPF file, a loaded TPF or SPF file is not generated.
Chapter 9
Prepare the OpenArray
™
Plate for PCR
Remove the OpenArray
™
Plate from the instrument
9
60
QuantStudio
™
12K Flex OpenArray
™
AccuFill
™
System User Guide