1
2
1
Section of the tip box that is full
2
Section of the tip box that is empty
2.
(Optional)
Click
Auto-fill tip boxes
.
The status of all section of the tips boxes is set to full.
3.
Select the first section of the sample plate that will be used to fill the OpenArray
™
Plate.
Select the first section of the sample plate if multiple OpenArray
™
Plates are filled during a run. The
software selects the total number of sections that correspond with the total number of OpenArray
™
Plates.
In the example below, section 5 was selected. The group of sections 5, 6, 7, and 8 is highlighted
by the software because four OpenArray
™
Plates are being filled.
The position box displays the color that corresponds to the section of the sample plate.
2
1
1
Sample plate section (section 5, 6, 7, and 8 are highlighted)
2
Corresponding OpenArray
™
Plates
4.
Remove the foil from the appropriate sections of the sample plate, then click the checkbox to
confirm.
Remove the foil only from the sections of the sample plate that are used to load a single
OpenArray
™
Plate.
Note:
Do not remove the foil from all the sections of the sample plate at once.
5.
Close the instrument door.
Chapter 7
Set up and start a run with the full run workflow
Verify the run setup and start the run—full run workflow
7
QuantStudio
™
12K Flex OpenArray
™
AccuFill
™
System User Guide
53