• Allow the instrument to perform a self-test if the run is being started after the software is launched
(see “Perform a system self-test”
A self-test is performed only if
Full Run
is selected after the software is launched. It is not
performed before each run.
• Prepare the materials in the QuantStudio
™
12K Flex OpenArray
™
Accessories Kit.
These materials are used to seal the OpenArray
™
Plates.
IMPORTANT!
OpenArray
™
Plates must be sealed immediately after loading. For more information,
see the user guide or application guide for your assay.
Configure the experiment design for the full run workflow
A sample plate file (CSV file) is
optional
. A TPF or SPF file
is
required. See “File formats”
.
Navigate to the
Full Run
screen.
If sample plate files and TPF or SPF files are imported, the default folder locations are set in the
preferences (see “Set the preferences”
1.
In the
Configure design
pane, in the
Experiment type
section, select an experiment type.
•
Gene expression
•
Genotyping
•
3-sample fixed panel
•
4-sample fixed panel
2.
In the
Plate format
section, select a format.
The values in the
Plate format
section depend on the experiment type that was selected in step 1.
3.
If the
Pipettor
section is displayed, select a type of pipette.
•
Fixed
•
Adjustable
The
Pipettor
section is only displayed if
Gene expression
or
Genotyping
was selected in step 1.
The number of OpenArray
™
Plates displayed in the
Add your OpenArray plate serial numbers
section depends on the selections made in the previous steps.
4.
In the
Add your OpenArray Plate serial numbers
section, click
Choose File
, navigate to the
location of the TPF or SPF file, then select the file.
Repeat for each TPF or SPF file.
5.
(Optional)
In the
Add your sample plates - optional
section, click
Choose File
, navigate to the
location of the CSV file, then select the file.
The file must be a CSV file for a 96-well format.
Chapter 7
Set up and start a run with the full run workflow
Configure the experiment design for the full run workflow
7
QuantStudio
™
12K Flex OpenArray
™
AccuFill
™
System User Guide