Figure 31 Dimensions of the typical layout of the QuantStudio
™
12K Flex OpenArray
™
AccuFill
™
System
1
Monitor and keyboard
2
Computer tower
3
Electrical outlet
4
OpenArray
™
AccuFill
™
Instrument
Component clearances required for installation and
maintenance
During installation and maintenance, it is necessary to access the back and sides of the product. If the
back of the product faces a wall, ensure that there is sufficient clearance on the bench to rotate the
product.
IMPORTANT!
For safety reasons, the power outlet for the product must be accessible at all times.
• The computer workspace allows for proper ergonomics during use.
• The workspace should comfortably fit the instrument and the computer.
• The location accommodates the dimensions and weights specified (“Components (unpackaged)”
• The location safely supports 100 kg (220 lbs).
• Allow at least 10 cm (4 in.) of cleared space next to the fan to allow air intake and exhaust.
• The instrument should be placed at least 1 m (1 yd) above the floor in order to protect against
water damage.
• The instrument should be placed at least 190 cm (75 in.) from the ceiling to allow the door to be
opened for service access.
Appendix C
Site requirements
Component clearances required for installation and maintenance