137
Thermo Fisher Scientific
Operation Manual HM550
• Put one or two drops into the space (fig. 132.2) of the cross roller
bearings.
• e specimen clamping should be in the lower position.
• Also slightly lubricate the horizontal cylinder guide behind the
specimen clamping.
Note:
For the examination and re-adjustment of the microtome a routine mainte-
nance should be performed by trained service technician once a year.
Cleaning and care of the microtome during routine work:
c
aution
A regular and/or immediate disinfection of the blade carrier, the cryo
chamber as well as all the other potentially contaminated parts of the
instrument is highly recommended. All commercially available cryostat
disinfection solutions can be used.
• Open the cleaning opening via the attached tool (fig. 134.1).
• For this, insert the tool (fig. 133.1) through one of the middle ven-
tilation slots.
• Turn the tool by 90° so that the t-shaped holder latches into the
adjacent slots.
• Pull the tool to open the grid (fig. 133.1) of the cleaning hole.
Figure 132.
Lubrication of the cross roller bearings.
Cleaning the cooling
lamella
Maintenance and care of the cryostat
Cleaning the cooling lamella
[1]
[3]
[2]
[4]