76
Thermo Fisher Scientific
Operation Manual HM550
Magnetic disposable
blade carrier MC
[2]
[1]
[4]
[3]
Figure 78.
Magnetic disposable blade carrier MC.
• Bring the clamping lever (fig. 77.3) into clamping position.
• e selected clearance angle is now fixed in its position.
Note:
Usable cuts are only achieved at a clearance angle of 10° or more!
e magnetic disposable blade carrier takes up ermo Scientific Microm
blades with the dimensions 80 x 19 mm and/or 60x19 mm and a facet
angle between 24° and 30°.
Inserting the blade
Note:
Special ermo Scientific magnetic blades must be used only!
• Insert the blade into the blade slider and put it onto the magnetic
surface (fig. 78.1) of the blade carrier from the side.
Note:
Always insert the blade in a way that the large facet of the blade is inserted
downwards. So it is invisible for the user.
• When inserting the blade, the blade slider together with the blade
must be press downwards, so that the blade is put properly against
ledge.
Operating instructions
Knife carriers