98
Thermo Fisher Scientific
Operation Manual HM550
• Lift the vacuum anti-roll hood (fig. 102.4), the flap (fig. 102.6)
beneath falls back in place and the slot gets closed.
e magnetic blade carrier is suitable to take up ermo Scientific
Microm blades with the dimensions 60 x 19 mm and a facet angle of
approx. 33°.
Magnetic blade
carrier MV
[7]
[5]
[4]
[1]
[10]
[2]
[3]
[6]
[8]
[9]
[11]
Figure 103.
Magnetic blade carrier MV.
Note:
Only special ermo Scientific magnetic blades must be used!!
• Insert the blade (fig. 103.1) into the blade slider (fig. 103.2) and
put it on the magnetic surface (fig. 103.2) of the blade carrier from
the side.
Note:
To move the blade to the side, press the blade slider downwards so that the
blade is properly put against the ledge (fig. 103.3) of the blade carrier.
• To move the blade to the side, always use the blade slider (fig.
104.2) and shift it.
If the cutting area of the blade is not longer usable, the blade can be
moved:
• Put the blade slider (fig. 104.2) onto the blade and shift it.
Inserting the blade
Operating instructions
Knife carrier for Vacutome