82
Thermo Fisher Scientific
Operation Manual HM550
• e selected operating mode is shown on the display as CONTIN,
SINGLE and INTERV and the corresponding LED in button (fig.
85.1) lights up.
• Press button (fig. 85.1) until the operating mode continuous stroke
CONTIN is shown on the display.
• To start a continuous cutting cycle, use the foot pedal or press the
two start buttons (fig. 85.3 and 85.4).
• To stop the continuous stroke, press foot pedal or one of the two
start buttons (fig. 85.3 or 85.4).
• However, if a cutting cycle has been started, it will continue through
and stop in the next lower reversal point.
Figure 84.
Display: Setting the operation mode.
OMVPD
Continuous stroke
[4]
[1]
[3]
Figure 85.
Keypad right.
Operating instructions
Option M – motorized cutting drive