Figure 104.
Insert a blade into the magnetic blade carrier MV.
Clearance angle adjustment
Note:
e blade slider must always be put on in a way that the inscription can be
read.
• is way, it is guaranteed that the slanted edge (fig. 104.5) of the
blade slider shows to the rear side of the blade.
c
aution
To avoid the danger of injury on the blade during adjustment of speci-
men, always position the knife guard (fig. 103.10) over the blade edge.
• Loosen the clamping lever (fig. 103.7).
• Swivel the upper part of the magnetic blade carrier on the base until
the desired clearance angle is adjusted.
• e set clearance angle can be read on the scale (fig. 103.8) on the
left side of the base.
• Again tighten the clamping lever.
• is way, the set clearance angle is fixed in its position.
Operating instructions
Knife carrier for Vacutome