Figure 129.
Removing the microtome.
Figure 130.
Removal of the heated window.
• Move the microtome forwards and remove it from the cryo-chamber.
• oroughly clean and wash the interior.
c
aution
Chloric cleaning agents must not be used for cleaning purposes. In case
of a malfunction hazardous gases might be generated together with the
cooling brine Tyfoxit of the cooling system.
• Pull off the red stopper at the bottom of the chamber. e cleaning
liquid is collected in the container (fig. 131.1) with a volume of 4,8
l at the front lower side
HM550M
VD
Maintenance and care of the cryostat
Removing the microtome