3
Description of the device
Workspace Components
3-22
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Shelf system
Figure 3-16.
Shelf system components
The support rails (3/
) of the shelf system are perforated every 42 mm (1.66 inch) so that
the support hooks (2/
) can be inserted variably for any culture container size required. The
shelves (1/
) have an integrated tilt protection and withdrawal stop.
S
TERI-CYCLE
i250
models
come equipped with a single-piece bottom insert, as shown on the left side in
S
TERI-CYCLE
i160
with a cutout for refilling (4/
). For details on using the shelf system, see
.
Note
Operating conditions:
When accessories are to be operated in the workspace of the CO
2
incubator, the ambient condition
requirements must be observed (see table). The energy introduced into the workspace affects the
beginning of the temperature control range. When additional heating sources are introduced into
the workspace, condensation (e.g. at the glass door) may occur.
Introduced energy
Beginning of the temperature control range
General
Example:
RT* = 21 °C
0 W
RT + 3 °C
24 °C
5 W
RT + 6.5 °C
27.5 °C
10 W
RT + 9.5 °C
30.5 °C
15 W
RT + 13 °C
34 °C
20 W
RT + 16 °C
37 °C
*RT = Room temperature
1
2
3
4
Summary of Contents for STERI-CYCLE i160
Page 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Page 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...