TERI-CYCLE
i160
/
i250
6-47
Figure 6-53.
Failure message Overtemperature
2. To exit the error display:
• Press the
End
key [3].
• The error message disappears.
• The temperature display field is displayed with a red bounding box.
3. Resetting the error message:
• Turn the device off.
4. Open the door and allow the workspace to cool down.
5. Turn the device on.
When overtemperature protection kicks back in although possible causes (see error log) have been
remedied, shut down the unit and call technical field service.
Measures after power outage
After a power outage, the formation of condensation can occur at the sensors due to moisture. This can
impair the sensor function and lead to erroneous values being displayed or to failure messages (sensor
breakage, see
).
The following measures must be taken to ensure safe operation:
1. Drain the water and wipe the interior dry.
2. Let the device warm up without water for 1 hour to 55 °C.
3. Then let the device cool down with the doors opened.
4. After this, start the device at incubation temperature according to Chapter
.
Alternatively, or when the above mentioned measures were not effective, start a disinfection run at 180
°C. Refer to Chapter
“Steri-run decontamination routine“
The disinfection routine can be interrupted after appro. 1 hour. The sensors should then have dried.
Summary of Contents for STERI-CYCLE i160
Page 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Page 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Page 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...