9
Maintenance
CO2 calibration procedure
9-6
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Figure 9-4.
Gas-tight screen measurement opening
CO
2
calibration procedure
Measurement example:
• CO
2
set value: 5 %
Reference value: 5.6 %
1. Press the CO
2
display key.
• The CO
2
menu is displayed.
2. To exit the CO
2
menu:
• Press the
End
key [3].
3. To enter the Calibration submenu:
• Press the
Calibration
key.
1