Activating the auto-start routine
Preparations for the start:
1. Make sure that the valves of the CO
2
/O
2
/N
2
gas supply system are open.
2. Fill the reservoir with purified water. Do not exceed the upper level mark.
Launching the auto-start routine:
1. Press the auto-start key.
• The auto-start instruction menu is displayed.
Figure 6-13.
Activating the auto-start routine
2. To exit from the auto-start instruction menu and stop auto-start:
• Press the
End
key [3].
3. Refill water, if necessary.
4. Activate the auto-start routine:
• Press the
Start
key.
5. To air the workspace, open both device doors. When the audible alarm sounds after 30 seconds,
Auto-start routine aborted due to error:
The auto-start routine is aborted if:
• an error is detected in the temperature control loop,
• an error is sensed in the CO
2
control circuit,
• the water filling level is insufficient,
• the CO
2
value to be set is out of tolerance.
Dry execution of auto-start routine:
If the routine is to be run dry (without water in the water reservoir of the workspace), the water level
sensor must be disabled prior to its start (chapter
).
Summary of Contents for STERI-CYCLE i160
Page 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Page 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Page 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...