6
Handling and control
User configuration
6-26
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Reminder interval setting
The reminder intervals are integral components of the alarm and monitoring system of the device
control. For the two essential functions steri-run and auto-start as well as for routine service work, the
user can set dates that trigger an alarm whenever they occur.
The counting begins at 00:00 hrs of the day on which the previously set reminder interval has elapsed.
On the due date, the display shows a reminder message for the activated reminder interval:
• Steri-run: Please run Steri-Run routine.
• auto-start: Please run auto-start routine. Appears upon successful completion of the Steri-Run
decontamination routine.
• Service interval: Request service. The service message can be confirmed. In this case, the Request
Service icon is displayed.
After the routines have been run successfully, the reminder messages are hidden.
Factory settings
Steri-run decontamination routine
90 days
auto-start routine
Off
Service interval
Off
HEPA interval
365 days
Reminder interval setting
1. Press the
Menu
key.
2. Select the Reminder intervals menu option.
• The selection dialog shown in
Figure 6-29.
Choose the desired reminder interval function.
3. Select the appropriate menu option, for example
Steri-run
.
Summary of Contents for STERI-CYCLE i160
Page 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Page 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Page 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...