2
Installation of the device
Room ventilation
2-2
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Room ventilation
When CO
2
-/O
2
-/N
2
is supplied, the workspace of the incubator is slightly pressurized. The pressure is
released through the pressure compensation opening into the operating room.
Pressure compensation and any opening of the glass door/gas-tight screen during the operation of the
device will release very small quantities of CO
2
-/O
2
-/N
2
into the operating room. The room
ventilation must be capable of carrying the released gas safely off into the open.
In addition, heat dissipating from the device during continuous operation may cause a change in the
room climate.
• Therefore, the
S
TERI-CYCLE
i160
/
S
TERI-CYCLE
i250
must only be installed in rooms with sufficient
ventilation.
• Do not install the device in room recesses without ventilation.
• The room ventilation should be a technical ventilation that complies with the national
requirements for laboritories or some other suited ventilation system with appropriate capacity.
Space requirements
Figure 2-1.
Device dimensions
WARNING
EMERGENCY STOP!
The mains socket must be accessible at all times for the case of an EMERGENCY STOP. The
mains socket must be easily identifiable by the user and freely accessible at all times.
Note
Accessibility of the devices:
To ensure the accessibility for care and maintenance works, keep larger side and rear distances.
STERI-CYCLE
i250
780
934
970
173
945
90
75
STERI-CYCLE
i160
637
≥173
90
7
5
0
9
All dimensions in mm.
Summary of Contents for STERI-CYCLE i160
Page 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Page 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Page 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...